GUI Reference
R&S
®
ZNB/ZNBT
674
User Manual 1173.9163.02 ─ 53
Furthermore, the user preset file is used as a template, whenever
are created from the GUI.
Remote command:
SYSTem:PRESet:USER:NAME
SYSTem:PRESet:USER[:STATe]
Start in Preset
If this option is active, the R&S
ZNB/ZNBT FW always starts with the configured
. Otherwise it starts with the previous state, i.e. the settings that were per-
sisted during the previous shutdown.
Remote command:
Calibration Tab
Provides general system error correction (calibration) settings.
Auto Power Reduction for Cal Unit
Sets the source power at all test ports to -10 dBm while an automatic calibration is
active. Applying this source power to the ports of the calibration unit ensures best
accuracy of the automatic calibration. The source powers are reset to their original val-
ues after the calibration is completed. The automatic power reduction can be deactiva-
ted in case that the test setup introduces a large attenuation.
Remote command:
SYSTem:COMMunicate:RDEVice:AKAL:PREDuction[:STATe]
Setup Softtool