SYSTem:SMATrix:OPTimization
<Unit>
Sets/gets the switch matrix route selection algorithm.
Parameters:
<Unit>
SPEed | PRECision
SPEed
–
Optimizes measurement speed by minimizing the total
number of switching procedures. This is highly recommended for
matrices with mechanical switches.
PRECision
–
Optimizes measurement precision by always using
best possible routes (according to matrix-specific route prioriza-
tion).
SYSTem:SHUTdown
[<Unit>]
Performs a shutdown or restart of the FW or OS.
If the optional parameter
<Unit>
is omitted, Windows is shutdown afer a time-out
period of 10 seconds.
Setting parameters:
<Unit>
HALT | REBoot | ABORt | CLOSe | RESTart
HALT
Windows is shutdown afer a time-out period of 20 seconds
REBoot
Windows is restarted afer a time-out period of 20 seconds
ABORt
Abort a Windows shutdown/restart. This can only be used during
the time-out period.
CLOSe
Close the firmware.
RESTart
Restart the firmware.
Example:
SYST:SHUT
Switch the analyzer to standby state.
Usage:
Setting only