Command Reference
R&S
®
ZNB/ZNBT
1118
User Manual 1173.9163.02 ─ 53
●
[SENSe:]CORRection:COLLect:CHANnels:MCTYpes
before executing this command. For background information see
lel Calibration of Multiple Channels"
Calibration data is acquired using
[SENSe<Ch>:]CORRection:COLLect[:
.
Suffix:
<Ch>
.
Channel number
Parameters:
<CalName>
Name of the calibration (string parameter). The name serves as
a reference to delete a particular set of system correction data
(
[SENSe<Ch>:]CORRection:COLLect:DELete
<CalType>
ARTosm | ARPTosm | REFL | RSHort | FOPort | FTRans |
RTRans | FRTRans | OPTPort | TOSM | UOSM | TRL | TOM |
TSM | TRM | TNA | PTOSm | PTOM | PTSM | PTRM | PTRL |
PTNA | PUOSm | VPUosm | MTOSm | MUOSm | MOSM | MTRL
Calibration types; see list below.
<TestPort1>
First calibrated test port number. For an n-port calibration type, n
port numbers must be specified. If more than n numbers are
defined, the spare numbers (the last ones in the list) are ignored.
Entering less than n numbers causes an error message. For a
one path two port calibration (
OPTPort
), the first port no.
denotes the node port (fully calibrated port).
<TestPort2>
Second calibrated port number.
<TestPort3 or
AdapterCalKit>
Either the third calibrated port number, or - for Adapter Removal
calibrations only - the file path (relative to the current directory)
where the analyzer shall store the adapter characterization
obtained during the calibration.
The latter is optional. If a file path is provided the characteriza-
tion file (in Touchstone s2p format) is generated when the cor-
rection terms are finally calculated (
]CORRection:COLLect:SAVE:SELected[:DUMMy]
<TestPorts>
More ports to be calibrated.
Example:
:SENSE1:CORRECTION:COLLECT:METHOD:DEFine
'Test', ARTosm, 1,2, 'Adapter.s2p'
Initiate an Adapter Removal calibration for ports 1 and 2 on
channel 1 and store the adapter characterization in
<MMEMORY:CDIR?>\Adapter.s2p
.
For another example see
Manual operation:
See
The supported calibration types are listed below.
SCPI Command Reference