Command Reference
R&S
®
ZNB/ZNBT
1234
User Manual 1173.9163.02 ─ 53
*RST:
RRAFter
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]
Manual operation:
See
"Flatness Cal – Pwr Cal Method"
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect:PMReadings
<NoReadings>
Selects the number of power meter readings for source power calibration method
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect:METHod RRAFter
.
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the
parameter is valid for all channels.
<PhyPt>
Calibrated port number. This parameter is ignored; the parame-
ter is valid for all sources.
Parameters:
<NoReadings>
Number of power meter readings. If the value exceeds the maxi-
mum number of calibration sweeps (
POWer<PhyPt>:CORRection:COLLect:AVERage[:COUNt]
the latter is adjusted.
Range:
1 to 100
*RST:
1
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]
.
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect:RRECeiver
<arg0>
Enables or disables calibration of the reference receiver together with the source
power calibration.
Note:
It is not possible to disable flatness calibration
(
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect:FLATness
) and the refer-
ence receiver calibration simultaneously.
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the verifi-
cation is either enabled or disabled for all channels.
<PhyPt>
Calibrated port number. This parameter is ignored; the verifica-
tion is either enabled or disabled for all sources.
Parameters:
<arg0>
Enables (ON) or disables (OFF) reference receiver calibration.
*RST:
ON
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect:
SCPI Command Reference