Firmware V3.12
R&S
®
ZNB/ZNBT
18
User Manual 1173.9163.02 ─ 53
–
Logarithmic interpolation
See "Interpolation" column in
Chapter 6.6.1.2, "Define Limit Lines Dialog"
●
Cal kit data added:
–
N
50
Ω
Keysight 85032B, 85032F, 85515A
R&S
ZN-Z170
typical
Spinner BN533843, BN533844, BN533863, BN533864
–
3.5
mm
R&S
ZN
‑
Z135
typical
–
2.92
mm
R&S
ZN-Z129
typical, R&S
ZN-Z129E
typical
–
7-16
Keysight 85038A
Spinner BN533845, BN533846, BN533865, BN533866
●
Faster ground loop de-/embedding
●
Cal kit standards: loss limit increased to 100,000 GΩ/s
●
Menu bar: optimized submenu structure for trace math ("Trace" > "Trace Config" >
"Trace Math")
●
Remote control: reduced number of
ACQuire
commands in Multiple Calibration
Types Mode
See the example in
[SENSe:]CORRection:COLLect:CHANnels:MCTYpes
Solved issues
●
Marker disappeared in zoom mode
●
Calibrated imbalance and CMRR traces did not show "Cal" label
●
Incorrect length offset calculation, if a Short was connected when the "Auto Length"
or "Auto Length and Loss" function was called
●
Discrete markers did not always snap to sweep points
●
Trace math on formatted data: for source data traces, an active "Min Hold" or "Max
Hold" was ignored
●
Loading an unsuitable
snp
file (e.g. a
s3p
file on a 2 port instrument), caused a
firmware exception and made the VNA stop sweeping
●
"Load Simulation Data" from s
n
p files with
n
> 9 did not work reliably
●
Marker values not displayed for Y and Z parameter traces in "dB Mag" format
●
Firmware simulation: when switching between simulated instruments, the fixture
de-/embedding tool support sometimes got confused about the number of available
ports