Command Reference
R&S
®
ZNB/ZNBT
863
User Manual 1173.9163.02 ─ 53
'Y-S11' | ... | 'Y-SSS11' | ... | 'Y-SCC11' | ... | 'Y-
SDD11' | 'Z-S11' | ... | 'Z-SSS11' | ... | 'Z-SCC11' | ...
| 'Z-SDD11' | ...
S-parameters converted to matched-circuit admittances and impedances
with port modes and port numbers like for normal mode S-parameters. For
transmission parameters, this refers to series resistances.
'SHUNT-S21'
Converted impedances measured using shunt-thru method. Use
to set the probe tip impedance.
'A1' | ... | 'A01' | ... | 'B1' | ... | 'B01' | ...
... | 'A1SAM' | 'A1AVG' | 'A1AMP'
Wave quantities A<meas>D1 (for a wave) and B<meas>D1.
The strings SAM, AVG, AMP appended to the wave quantities denote a nor-
mal (sample, SAM), AVG Real Imag (AVG), or AVG Mag Phase (AMP)
detector. The observation time for average detectors is set via
[SENSe<Ch>:]SWEep:DETector:TIME
.
'A1D2' | ... | 'A01D02' | ... | 'B1D2' | ... | 'B01D02' | ...
... | 'A1D1SAM' | 'A1D1AVG' | 'A1D1AMP' | ...
Wave quantities A<meas>D<drive> (for a wave) and B<meas>D<drive> (for
b wave) for logical ports <meas> and <drive>.
The strings SAM, AVG, AMP appended to the wave quantities denote a nor-
mal (sample, SAM), AVG Real Imag (AVG), or AVG Mag Phase (AMP)
detector.
'A1G2' | ... | 'A01G02' | ... | 'B1G2' | ... | 'B01G02' | ...
| 'A1G1SAM' | 'A1G1AVG' | 'A1G1AMP' | ...
Wave quantities with an external generator providing the stimulus signal
(G<no> for generator no.). The strings SAM, AVG, AMP appended to the
wave quantities denote a normal (sample, SAM), AVG Real Imag (AVG), or
AVG Mag Phase (AMP) detector.
'B2/A1' | ... 'B02/A01' | ...
... | 'B2/A1SAM' | 'B2/A1AVG' | B2/A1AMP' | ...
Ratio of wave quantities with port numbers like for normal mode S-parame-
ters.
The strings SAM, AVG, AMP appended to the wave quantities denote a nor-
mal (sample, SAM), AVG Real Imag (AVG), or AVG Mag Phase (AMP)
detector.
'B2D1/A1D1' | ... 'B02D01/A01D01' | ...
... | 'B2D1/A1D1SAM' | 'B2D1/A1D1AVG' 'B2D1/
A1D1AMP' | ...
Ratios of wave quantities with drive ports
The strings SAM, AVG, AMP appended to the wave quantities denote a nor-
mal (sample, SAM), AVG Real Imag (AVG), or AVG Mag Phase (AMP)
detector.
'B2G1/A1G1' | ... 'B02G01/A01G01' | ... 'B2G1/
A1G1SAM' | 'B2G1/A1G1AVG' 'B2G1/A1G1AMP'
| ...
Ratios of wave quantities with port numbers and external generator providing
the stimulus signal (G<no> for generator no.). The strings SAM, AVG, AMP
appended to the wave quantities denote a normal (sample, SAM), AVG Real
Imag (AVG), or AVG Mag Phase (AMP) detector.
'IMB21' | 'IMB12' | 'IMB31' | ...
Imbalance parameter Imb<receive_port><drive_port> for logical port num-
bers <receive_port> and <drive_port>. The logical ports must be different
and at least one of them must be balanced.
'CMRR11' | 'CMRR21' | 'CMRR12' | 'CMRR22' | ...
Common mode rejection ratio CMRR<receive_port><drive_port> parameter
for logical port numbers <receive_port> and <drive_port>, at least one of
them balanced.
'IMB1-23' | 'IMB23-1' | 'IMB1-24' | ...
Differential Imbalance parameters between a balanced and two single-ended
logical ports
'KFAC21' | 'KFAC12' | ...
Stability factor K (for unbalanced ports only)
'MUF121' | 'MUF112' | ...
Stability factor 1 (for unbalanced ports only)
'MUF221' | 'MUF212' | ...
Stability factor 2 (for unbalanced ports only)
'Pmtr1G1' | 'Pmtr2G1' | 'Pmtr3D1' | ...
Power sensor measurement using a power meter 'Pmtr<no>' and either an
external generator 'G<no>' or an analyzer source port 'D<no>'
'DC1D1' ... 'DC4D1' | 'DC1D01' ... 'DC4D01'
'DC1V' | 'DC10V'
DC measurement DC 1 to DC 4. D1 or D01 etc. denotes the (logical) drive
port
For R&S ZVAB compatibility; query returns 'DC1D1' or 'DC2D1')
SCPI Command Reference