Command Reference
R&S
®
ZNB/ZNBT
1247
User Manual 1173.9163.02 ─ 53
SOURce<Ch>:POWer<PhyPt>[:LEVel][:IMMediate]:SLOPe
<Slope>
Defines a linear factor to modify the internal source power at port
<PhyPt>
as a func-
tion of the stimulus frequency.
Suffix:
<Ch>
.
Channel number
<PhyPt>
Test port number of the analyzer
Parameters:
<Slope>
Port-specific slope factor
Range:
-40 dB/GHz to +40 dB/GHz. The resulting power
range over the entire frequency sweep must be
within the power range of the analyzer; refer to the
data sheet.
Increment: 0.1-dB/GHz
*RST:
0 dB/GHz
Default unit: DB/GHZ
Example:
See
SOURce<Ch>:POWer<PhyPt>[:LEVel][:
Manual operation:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:GENerator<Gen>:LEVel:OFFSet
<Offset>
Specifies a gain (positive values) or an attenuation (negative values) in the signal path
between the external generator and the calibrated reference plane. The value has no
impact on the generator power.
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt>
Calibrated port number. This suffix is ignored; the generator is
selected via
<Gen>
.
<Gen>
Generator number
Parameters:
<Offset>
Gain or attenuation value
Range:
-300 dB to +300 dB (adjust to the test setup)
Increment: 0.01 dB (other values are rounded)
*RST:
0 dB
Default unit: dB
SCPI Command Reference