Command Reference
R&S
®
ZNB/ZNBT
1099
User Manual 1173.9163.02 ─ 53
[SENSe<Ch>:]CORRection[:STATe]
<Boolean>
Enables or disables the system error correction for channel
<Ch>
.
Suffix:
<Ch>
.
Calibrated channel number
Parameters:
<Boolean>
Enables (ON) or disables (OFF) the correction.
*RST:
ON
Example:
*RST; :CORR?
Reset the instrument and query whether channel 1 is system
error corrected. The response is
1
.
Manual operation:
See
[SENSe<Ch>:]CORRection:COLLect[:ACQuire]:SELected
<Type>, <TestPort>[,
<SecondPortOrAdapter>, <Dispersion>, <DelayTimePhase>]
Starts the acquisition of measurement data for the selected standard and port(s). The
standards are reflection or transmission standards and can be connected to arbitrary
analyzer ports.
Note:
●
The calibration measurement has a variable timeout:
Timeout = (Sweep time /
Number of sweep points) * 3 + 0.1 s
The timeout may be reached e.g. while the measurement waits for the specified
trigger event. It is used for all standard measurements including power calibration
sweeps.
●
For a sliding match, the R&S
ZNB/ZNBT can acquire measurement data for up to
20 positions per port. Multiple calls of
[SENSe<Ch>:]CORRection:COLLect[:ACQuire]:SELected SLIDe,
<TestPort>
for the same test port, implicitly increases the position until position
20 has been recorded. Subsequent calls will start over at position 1, overwriting the
previously acquired data.
Suffix:
<Ch>
.
Channel number of the calibrated channel
SCPI Command Reference