GUI Reference
R&S
®
ZNB/ZNBT
461
User Manual 1173.9163.02 ─ 53
2.
"Connections":
Define the port assignments between the R&S
ZNB/ZNBT and
the calibration units.
3.
"SMARTerCal":
a) For each configured SMARTerCal and port assignment, acquire error correc-
tion data for the required calibration standards (provided by the selected cali-
bration unit).
b) Apply the resulting error terms to the active channel.
c) Acquire the power calibration data at the selected source ports.
d) Use this result to correct the absolute receiver powers at all calibrated ports.
Background and related information
●
The "Multiple Cal in Calibration Wizard" feature is disabled by default and has to be
activated in the
of the "System Config" dialog.
●
to perform an additional source flatness calibration (see
Chapter 5.5.7.3, "Combining SMARTerCal with Scalar Power Calibration"
●
Chapter 6.11.1.8, "SMARTerCal Wizard (Cal Unit)"
on page 451 for fur-
ther information.
Step 1: Ports
The first page of the wizard presents the list of already configured calibrations and
allows you to:
●
"Add" new calibrations
●
"Delete" or "Modify" existing calibrations
Add
Adds a new system error correction.
The "Add" button opens the
without pre-selected ports.
Delete
Deletes the selected calibration from the list of configured calibrations.
Cal Softtool