
Command Reference
R&S
®
ZNB/ZNBT
1138
User Manual 1173.9163.02 ─ 53
See
Chapter 5.5.5.4, "Multiple Port Assignments"
on page 172 for background infor-
mation.
Note
:
Use
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:COUNt?
to get the
number of port assignments of the "current" calibration.
Suffix:
<Ch>
.
Channel number
Example:
See
Chapter 9.2.5.2, "MultiCal (with Calibration Unit)"
Usage:
Query only
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:COUNt?
Returns the number of port assignments of the "current" calibration, i.e. the calibration
last created using
[SENSe<Ch>:]CORRection:COLLect:AUTO:CONFigure
This is particulary useful if the default port assignments were established using
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:DEFine:DEFault
.
See
Chapter 5.5.5.4, "Multiple Port Assignments"
on page 172 for background infor-
mation.
Note
:
In MultiCal scenarios, use
[SENSe<Ch>:]CORRection:COLLect:AUTO:
to get the number of port assignments of all calibrations.
Suffix:
<Ch>
.
Channel number
Example:
See
[SENSe<Ch>:]CORRection:COLLect:AUTO:
Usage:
Query only
Manual operation:
See
[SENSe<Ch>:]CORRection:COLLect:AUTO:ASSignment:DEFine:DEFault
<TestPort1>, <TestPort2>...
Creates the default port assignment(s) for the specified test ports.
The number of created assignments can be queried using the
]CORRection:COLLect:AUTO:ASSignment:COUNt?
command.
Note
that during the corresponding calibration sweep the R&S
ZNB/ZNBT expects the
physical port connections to be established exactly as specified by the created port
assignments.
SCPI Command Reference