Command Reference
R&S
®
ZNB/ZNBT
1172
User Manual 1173.9163.02 ─ 53
Suffix:
<Ch>
.
Channel number
Parameters:
<Mode>
AUTO | LDIStortion | LNOise
AUTO
automatic/adaptive gain control (AGC) according to the RF input
level
LDIStortion
fixed, small IF gain, for high input levels
LNOise
fixed, large IF gain, for low input levels
*RST:
LDIStortion
Parameters for setting and query:
<ReceiverName>
String parameter defining the wave(s) followed by the drive port;
see examples below and
Example:
*RST; :SENSe:POWer:GAINcontrol:GLOBal MANual
Activate independent GC settings for each drive port.
SENSe:POWer:GAINcontrol 'B2D1', LNO
Set the GC for the received wave b
2
(port 2) to "Low Noise". The
setting applies while port 1 is the drive port.
SENSe:POWer:GAINcontrol 'A2B2B3D2', LNO
Set the GC mode for the waves a
2
, b
2
, and b
3
to "Low Noise".
The setting applies while port 2 is the drive port.
SENSe:POWer:GAINcontrol? 'B1D2'
Query the GC setting for the wave b
1
while port 2 is the drive
port. The response is
1, LDIS
(default setting).
Manual operation:
See
"Drive-port specific settings"
[SENSe<Ch>:]POWer:GAINcontrol:ALL
<Mode>
Applies the same manual gain control (GC) <Mode> to all a and b wave receivers.
Note:
●
Without the
option only the gain of the measurement
receivers (the b waves) can be statically set (to LNOise or LDIStortion); the refer-
ence receiver always uses automatic gain control (AUTO). With the option availa-
ble, also the gain of the reference receivers (the a waves) can be set statically.
●
If you also want to distinguish between different sweep segments, use
[SENSe<Ch>:]SEGMent<Seg>:POWer:GAINcontrol:ALL
Suffix:
<Ch>
.
Channel number
Setting parameters:
<Mode>
AUTO | LDIStortion | LNOise
SCPI Command Reference