Command Reference
R&S
®
ZNB/ZNBT
1096
User Manual 1173.9163.02 ─ 53
Parameters:
<Boolean>
ON
–
calibrate all channels.
OFF
–
calibrate the active channel only.
Example:
See
[SENSe<Ch>:]CORRection:COLLect:SAVE:
Manual operation:
See
[SENSe:]CORRection:COLLect:CHANnels:MCTYpes
<Boolean>
Toggles the
M
ultiple
C
alibration
TYpes
mode that allows you to calibrate a subset of
the available channels using channel-specific ports and calibration types (see
ter 5.5.8, "Parallel Calibration of Multiple Channels"
Enable the MCTYpes mode before defining the calibrations to be performed using
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine
. Disable the MCTYpes
mode after the calibrations were saved using
. The latter must be called only
once
, with <Ch>
referring to one of the channels to be calibrated (see the example below).
Parameters:
<Boolean>
SCPI Command Reference