Command Reference
R&S
®
ZNB/ZNBT
1186
User Manual 1173.9163.02 ─ 53
Example:
SEGM1:INS:SEL DWEL
Select the meas. delay to determine the sweep time in a new
sweep segment no. 1.
SEGM1:INS 1MHZ, 1.5MHZ, 111, -21DBM, 0.01S, 0,
10KHZ
Create a sweep segment with a sweep range between 1.0 MHz
and 1.5 MHz and a meas. delay of 10 ms.
SEGM1:SWE:TIME?
Query the sweep time in the new segment.
Manual operation:
See
[SENSe<Ch>:]SEGMent<Seg>:POWer[:LEVel]
<IntSourcePow>
Defines the power of the internal signal source in sweep segment no.
<Seg>
. At the
same time, the command activates separate power control in all sweep segments
(
[SENSe<Ch>:]SEGMent<Seg>:POWer[:LEVel]:CONTrol
).
Suffix:
<Ch>
.
Channel number
<Seg>
Sweep segment number
Parameters:
<IntSourcePow>
Internal source power
Range:
-300 dBm to +200 dBm. The usable power range is
frequency-dependent; refer to the data sheet.
*RST:
-10 dBm
Default unit: dBm
Example:
SEGM:ADD
Create a new sweep segment no. 1 in channel no. 1 using
default settings and thus -10 dBm internal source power.
SEGM:POW -20
Decrease the power to -20 dBm.
Manual operation:
See
[SENSe<Ch>:]SEGMent<Seg>:PORT<PortId>:POWer[:LEVel]
<IntSourcePortPow>
Defines a port-specific power level for port <PortId> in sweep segment <Seg> of chan-
nel <Ch>.
Requires that a port-specific power level has been previously enabled using
[SENSe<Ch>:]SEGMent<Seg>:PORT<PortId>[:STATe]
.
Suffix:
<Ch>
.
Channel number
<Seg>
Sweep segment number
<PortId>
Port number
SCPI Command Reference