User Manual 1173.9163.02 ─ 53
CALCulate<Chn>:TTIMe:STATe
<Boolean>
Enables/disables the
Note:
The rise time measurement can only be enabled if the active trace is real
REAL
), Time Domain is enabled (
ON
), and the Low Pass Step time domain transform is used
CALCulate<Chn>:TRANsform:TIME[:TYPE]
LPASs
and
STEP
). The latter, in turn, requires the stimulus grid to
be harmonic. This can be achieved, for example, using
ON
.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean>
ON – Rise time measurement enabled
OFF – Rise time measurement disabled
*RST:
OFF
Options:
R&S
ZNB/ZNBT-K20
Manual operation:
See
CALCulate<Chn>:TTIMe:THReshold
<ThresholdEnum|LowerThreshold>[,
<UpperThreshold>]
Defines the lower/upper threshold for the rise time measurement.
The thresholds can either be specified by enum constants for the standard 10–90% or
20–80% rise times, or as integer percentages.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<ThresholdEnum|LowerThreshold>
T1_9: 10–90% rise time
T2_8: 20–80% rise time
<integer>: Lower rise time threshold as integer percentage
*RST:
T1_9:
<UpperThreshold>
Upper rise time threshold as integer percentage
Options:
R&S
ZNB/ZNBT-K20
Manual operation:
See
8.3.2
CONFigure Commands
The
CONFigure...
commands create and delete channels or traces and assign
channel and trace names. The commands are device-specific.
SCPI Command Reference