
Command Reference
R&S
®
ZNB/ZNBT
1251
User Manual 1173.9163.02 ─ 53
SOURce<Ch>:POWer<PhyPt>:PERManent[:STATe]
<Boolean>
Defines whether the source power is permanently on ("generator port").
For the R&S
ZNBT, among the analyzer ports driven by the same internal source, more
than one can operate as generator port. However, all these ports must have identical
arbitrary frequency and power settings. If
SOURce<Ch>:POWer<PhyPt>:PERManent:STATe
is set to
ON
, then for channel
<Ch>
the arbitrary frequency and power settings of port
<PhyPt>
are copied all ports
driven by the same internal source.
Note
:
If
are part of the RF connection configuration, operation with
more than one internal source is
not
supported. In this case only external generators
can be used as permanent signal sources.
Suffix:
<Ch>
.
Channel number
<PhyPt>
Test port number of the analyzer
Parameters:
<Boolean>
ON - power at port
<PhyPt>
is permanently on.
OFF - power is only on for the partial measurements that require
the port as a drive port.
*RST:
OFF
Example:
SOUR:POW2:STAT ON
Switch the RF power at port no. 2 on.
SOUR:POW2:PERM ON
Activate a permanent signal source.
Manual operation:
See
SOURce<Ch>:POWer<PhyPt>:REDuce[:STATe]
<ReduceAtSweepEnd>
Deprecated
; superseded by
Enables/disables power reduction at sweep end.
This is a global setting.
Suffix:
<Ch>
.
Channel number. This suffix is ignored; the setting is valid for all
channels.
<PhyPt>
Physical port number. This suffix is ignored; the setting is valid
for all ports.
SCPI Command Reference