View / Modify Cal Kit Standards Dialog
The "View / Modify Cal Kit Standards" dialog shows the circuit model for a selected cal-
ibration standard. It is also used to define or edit the circuit model (offset and load)
parameters for a user-defined standard.
Related information
Refer to the following sections:
●
Chapter 5.5.2.4, "Cal Kit Files"
●
●
Chapter 5.5.2.1, "Calibration Standard Types"
Access:
> "View / Modify..."
The diagram in the "View / Modify Cal Kit Standards" dialog depends on the standard
type for which the dialog was opened. Moreover, it is possible to modify the circuit
model using the buttons in the upper right of the dialog.
Offset Parameters
The entries in the upper part of the "View / Modify Cal Kit Standards" dialog specify the
offset parameters for the transmission lines of the selected calibration standard.
The offset parameters depend on whether the circuit model is defined as "Agilent
Model" (see
Chapter 6.11.2.1, "Cal Connector Types Dialog"
●
In an "Agilent Model", a calibration standard is characterized by its "Delay" (in s),
its characteristic impedance "Z0" (in Ω) and its "Offset Loss" (in GΩ/s).