Command Reference
R&S
®
ZNB/ZNBT
1102
User Manual 1173.9163.02 ─ 53
–
A VNA cal kit file name without path but prefixed with "SD:"
refers to a specific cal kit file stored on the SD card inserted at
the active calibration unit.
–
A VNA cal kit file name
*.calkit
with path refers to a spe-
cific cal kit file stored in an arbitrary directory on the analyzer.
<TestPort>
<TestPort>
Test port numbers. For an n-port automatic calibration, n arbi-
trary (not necessarily consecutive) port numbers must be speci-
fied. The analyzer automatically detects the calibration unit ports
connected to each analyzer port.
Example:
CORR:COLL:AUTO '', 1, 2, 4
Perform an automatic 3-port calibration at test ports 1, 2, and 4
using the calibration unit's default calibration kit file and arbitrary
test ports of the cal unit.
Usage:
Setting only
Manual operation:
See
[SENSe<Ch>:]CORRection:COLLect:AUTO:CONFigure
<CalType>,
<Characterization>
Selects a calibration type and a cal unit characterization (cal kit file) for an automatic
calibration.
Suffix:
<Ch>
.
Channel number of the calibrated channel
Parameters:
<CalType>
FNPort | TOSM | FOPort | OPTPort | FRTRans | FTRans |
RTRans | REFL | RSHort | PFNPort
Calibration Type
FNPort
Full n-port (UOSM) calibration
TOSM
Full n-port calibration with characterized Through.
If the selected cal unit characterization does not contain a
Through characterization, the command silently falls back to a
FNPort (UOSM) calibration.
FOPort
Full one-port ("Refl OSM") calibration
OPTPort
"One Path Two Ports" calibration
If the selected cal unit characterization does not contain a
Through characterization, the command will fail.
SCPI Command Reference