Command Reference
R&S
®
ZNB/ZNBT
1238
User Manual 1173.9163.02 ─ 53
SOURce<Ch>:POWer<PhyPt>:CORRection:PPOWer
<RecCalPower>
Defines the source power which the R&S
ZNB/ZNBT uses to perform the first calibra-
tion sweep of the source power calibration ("Reference Receiver Cal Power"). The
power value is ignored if the R&S
ZNB/ZNBT is set to use the port power result
SOURce<Ch>:POWer<PhyPt>:CORRection:PSELect
CPOW
).
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the
parameter is valid for all channels.
<PhyPt>
Calibrated port number. This parameter is ignored; the parame-
ter is valid for all sources.
Parameters:
<RecCalPower>
Reference receiver cal power
Range:
Depending on source power range of the analyzer
and the power range of the external power meter.
*RST:
0 dBm
Default unit: dBm
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]
.
Manual operation:
See
"Reference Receiver Cal Power"
SOURce<Ch>:POWer<PhyPt>:CORRection:PSELect
<CalPower>
Qualifies how to define the source power which the R&S
ZNB/ZNBT uses to perform
the first calibration sweep of the source power calibration ("Reference Receiver Cal
Power").
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the
parameter is valid for all channels.
<PhyPt>
Calibrated port number. This parameter is ignored; the parame-
ter is valid for all sources.
Parameters:
<CalPower>
CPOWer | PPOWer
PPOWer
–
use cal power value defined via
POWer<PhyPt>:CORRection:PPOWer
CPOWer
–
use port power value
*RST:
PPOWer
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]
.
Manual operation:
See
"Reference Receiver Cal Power"
SCPI Command Reference