Command Reference
R&S
®
ZNB/ZNBT
1237
User Manual 1173.9163.02 ─ 53
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:
Usage:
Event
Manual operation:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:NREadings
<NoCalSweeps>
Sets a limit for the number of calibration sweeps in the source power calibration. The
command is valid for all channels, ports and external generators.
Equivalent command:
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect:
.
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the limit
applies to all sources.
<PhyPt>
Calibrated port number. This parameter is ignored; the limit
applies to all sources.
Parameters:
<NoCalSweeps>
Number of readings
Range:
1 to 100
*RST:
2
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]
Manual operation:
See
"Flatness Cal – Max Iterations"
SOURce<Ch>:POWer<PhyPt>:CORRection:OSOurces[:STATe]
<Boolean>
Switches off all other sources during the calibration sweep for channel
<Ch>
.
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt>
Calibrated port number. This parameter is ignored; the setting is
valid for all sources.
Parameters:
<Boolean>
ON - other sources not necessarily switched off.
OFF - other sources switched off during the calibration sweep.
*RST:
ON
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]
Manual operation:
See
SCPI Command Reference