[SENSe<Ch>:]FREQuency:CONVersion:MIXer:LOPort<Stg>
<SourceType>[,
<SourceNumber>]
Selects an analyzer or an external generator port as a signal source for the LO1 or
LO2 signal.
Note:
If
are part of the RF connection configuration, operation with
more than one internal source is
not
supported. In this case only external generators
can be used as local oscillators.
Suffix:
<Ch>
.
Channel number
<Stg>
Mixer stage (signals LO 1 or LO 2; see
]FREQuency:CONVersion:MIXer:STAGes
Parameters:
<SourceType>
PORT | GENerator | NONE
NONE
–
LO signal is not controlled by the analyzer.
PORT
–
analyzer port
GENerator
–
external generator
*RST:
NONE
<SourceNumber>
Number of analyzer port or generator.
Example:
See
[SENSe<Ch>:]FREQuency:CONVersion:MIXer:
Manual operation:
See
[SENSe<Ch>:]FREQuency:CONVersion:MIXer:MFFixed
<InputOutput>[,
<FixedFrequency>]
Assigns a fixed frequency to the RF, LO 1, LO 2, or to the IF signal. The fixed fre-
quency setting becomes active if the port is selected as a port with fixed frequency
(
[SENSe<Ch>:]FREQuency:CONVersion:MIXer:FIXed<Stg>
).
Suffix:
<Ch>
.
Channel number
Parameters:
<InputOutput>
RF | LO | LO1 | LO2 | IF
Fixed frequency for the specified port
RF – mixer input signal
LO | LO1 – local oscillator signal no. 1
LO2 – local oscillator signal no. 2, for 2-stage mixer measure-
ments (
[SENSe<Ch>:]FREQuency:CONVersion:MIXer:
IF – mixer output signal
Range:
Depending on the instrument model [Hz]. The incre-
ment (parameters UP or DOWN) is 0.1 kHz.
SCPI Command Reference