Command Reference
R&S
®
ZNB/ZNBT
1307
User Manual 1173.9163.02 ─ 53
The trigger duration must be shorter than the selected sweep interval (see
CHANnel<Ch>:AUXiliary<n>:INTerval
).
If the trigger duration is too long, you can modify one of the following settings:
●
Reduce the trigger duration, if possible.
●
Slow down the sweep: select a smaller bandwidth or increase the "Sweep Time"
setting.
Suffix:
<Ch>
.
Channel number
<n>
Auxiliary channel number. This suffix is ignored; it can be set to
1 or omitted.
Parameters:
<TrigOutDuration>
Pulse width between 1
μs (1E-6
s) and 1
s.
*RST:
1 us
Default unit: s
Example:
See
TRIGger:CHANnel<Ch>:AUXiliary<n>[:ENABle]
TRIGger:CHANnel<Ch>:AUXiliary<n>:INTerval
<Type>
Specifies whether the trigger output pulses are sent once every measurement point or
once every sweep.
Suffix:
<Ch>
.
Channel number
<n>
Auxiliary channel number. This suffix is ignored; it can be set to
1 or omitted.
Parameters:
<Type>
POINt | SWEep
ACQuisition
A trigger out signal is generated for every data acquisition phase
*RST:
SWEep
Example:
See
TRIGger:CHANnel<Ch>:AUXiliary<n>[:ENABle]
TRIGger:CHANnel<Ch>:AUXiliary<n>:OPOLarity
<Type>
Selects the polarity of the output trigger pulses; see
Suffix:
<Ch>
.
Channel number
<n>
Auxiliary channel number. This suffix is ignored; it can be set to
1 or omitted.
SCPI Command Reference