GUI Reference
R&S
®
ZNB/ZNBT
386
User Manual 1173.9163.02 ─ 53
CW Frequency
Sets the fixed frequency for
,
, and
The "CW Frequency" is also used as the channel base frequency for frequency-con-
verting measurements; see
Chapter 5.7.4, "Frequency Conversion Measurements"
Remote command:
[SENSe<Ch>:]FREQuency:FIXed
[SENSe<Ch>:]FREQuency[:CW]
SOURce<Ch>:FREQuency<PhyPt>:FIXed
SOURce<Ch>:FREQuency<PhyPt>[:CW]
6.8.2
Power Tab
The "Power" tab provides settings related to transmit and receive power. It is also dis-
played on the "Power Bw Avg" softtool.
6.8.2.1
Controls on the Power Tab
left = R&S
ZNB with
and 4 physical ports
right = R&S
ZNB/ZNBT with
and > 4 physical ports on the VNA and connected
switch matrices
If the number of test ports (on the VNA and connected switch matrices) is higher than
4, the "Step Attenuators" input fields are hidden. They are replaced by a "Step Attenua-
tors..." button that opens the following dialog:
Power
Determines the output power at the test ports for the sweep types "Lin Freq", "Log
Freq", "CW Mode", and "Time". Also determines the default output power for "Segmen-
ted" sweeps, where the output power can be set per segment.
Stimulus Softtool