![Rohde & Schwarz 1311.6010.22 User Manual Download Page 864](http://html.mh-extra.com/html/rohde-and-schwarz/1311-6010-22/1311-6010-22_user-manual_1477768864.webp)
Command Reference
R&S
®
ZNB/ZNBT
864
User Manual 1173.9163.02 ─ 53
'PAE21' | PAE12' | ...
Power added efficiency (referring to logical ports)
'IM3UO' | 'IM3LO' | 'IM3MO' | 'IM5UO' ... 'IM9MO'
Intermodulation product IM<order><side><at DUT> where <order> = 3 | 5 | 7
| 9, <side> = U | L | M (for upper or lower or major), at DUT output
'IM3UOR' | 'IM3LOR' | 'IM3MOR' | 'IM5UOR' ...
'IM9MOR'
Intermodulation product (as explained above) displayed in dB units
r
elative
to the measured lower tone level at DUT output ("intermodulation suppres-
sion")
'IP3UO' | 'IP3LO' | 'IP3MO' | 'IP5UO' ... 'IP9MO'
Intercept point IP<order><side>O where <order> = 3 | 5 | 7 | 9, <side> = U |
L | M (for upper or lower or major), at DUT output
UTI | UTO | LTI | LTO
Upper or lower tone at DUT input or output
*) Selecting a parameter Y...<n><m> or Z...<n><m> sets the range of port numbers to
be considered for the Y and Z-parameter measurement to <n>:<m>.
CALCulate<Ch>:PARameter:SDEFine:SENDed
<TraceName>, <Result>
Creates a trace and assigns a channel number, a name and a measurement parame-
ter to it. The trace becomes the active trace in the channel but is not displayed.
Similar to
CALCulate<Ch>:PARameter:SDEFine
, but in presence of balanced ports
this command distinguishes between
●
single-ended S-parameters for
logical
ports (
SSS<Li><Lj>
)
●
"raw" single-ended S-parameters referring to
physical
ports (
S<Pi><Pj>
)
Suffix:
<Ch>
.
Channel number
Setting parameters:
<TraceName>
Trace name
<Result>
Measured quantity
Usage:
Setting only
Manual operation:
See
CALCulate<Ch>:PARameter:SELect
<TraceName>
Selects an existing trace as the active trace of the channel. All trace commands with-
out explicit reference to the trace name act on the active trace (e.g.
).
CALCulate<Ch>:PARameter:SELect
is also neces-
sary if the active trace of a channel has been deleted.
Suffix:
<Ch>
.
Channel number.
Parameters:
<TraceName>
Trace name, e.g. 'Trc4'. See "Rules for trace names" in
SCPI Command Reference