GUI Reference
R&S
®
ZNB/ZNBT
462
User Manual 1173.9163.02 ─ 53
Modify
Edits the selected system error correction: opens the
with the
corresponding ports and calibration type pre-selected.
Next
.
Inactive as long as no SMARTerCal is defined.
Define Calibration Dialog
In MultiCal mode, the "Define Calibration" dialog allows you to define the individual
SMARTerCals: select the test ports to be calibrated, the calibration type, the calibration
unit and characterization, the power meter and power port.
Access:
SMARTerCal Wizard (Cal Unit, MultiCal)
> Step 1: "Ports" > "Add..." / "Mod-
ify..."
●
The "Multiple Cal in Calibration Wizard" (MultiCal) feature is disabled by default
and has to be activated in the
of the "System Config" dialog.
●
For background information on calibration types, see
The layout of the main panel depends on the number of test ports available.
Figure 6-28: Define Calibration dialog (MultiCal, > 4 test ports)
Ports
Selects the test ports to be calibrated.
It is possible to select any combination of two or more test ports. If you are only inter-
ested in a single port
p
, perform a two-port SMARTerCal for a port pair including
p
(as
Power Port).
Note:
Calibration and port de-/activation.
Cal Softtool