Command Reference
R&S
®
ZNB/ZNBT
1268
User Manual 1173.9163.02 ─ 53
Manual operation:
See
SYSTem:COMMunicate:RDEVice:GENerator<Gen>:SEPMode
<Type>
Defines the external generator power control after the sweep end.
Suffix:
<Gen>
.
Generator number
Parameters:
<Type>
OFF | LOW | KEEP | USER
OFF
–
power is switched off
LOW
–
switch to min. power
KEEP
–
keep power at the value selected via
POWer<PhyPt>:GENerator<Gen>:OFFSet
USER
–
set to end power selected via
RDEVice:GENerator<Gen>:SEPower
Example:
See
SYSTem:COMMunicate:RDEVice:GENerator<Gen>:
Manual operation:
See
SYSTem:COMMunicate:RDEVice:GENerator<Gen>:SEPower
<Power>
Defines the external generator power value after the end of a sweep. The setting is rel-
evant if the generator is switched to a user-selected power; see example.
Suffix:
<Gen>
.
Generator number
Parameters:
<Power>
End power
Range:
-145 dBm to +30 dBm
*RST:
0 dBm
Default unit: dBm
Example:
SYSTem:COMMunicate:RDEVice:GENerator:SEPMode
USER
Switch the generator to a user-selected power value after the
sweep is terminated.
SYSTem:COMMunicate:RDEVice:GENerator:SEPower
-10
Define a end power of -10 dBm.
Manual operation:
See
SYSTem:COMMunicate:RDEVice:PMETer:CATalog?
Queries the numbers of all configured external power meters. The response is a string
containing a comma-separated list of power meter numbers.
SCPI Command Reference