![Rohde & Schwarz 1311.6010.22 User Manual Download Page 617](http://html.mh-extra.com/html/rohde-and-schwarz/1311-6010-22/1311-6010-22_user-manual_1477768617.webp)
GUI Reference
R&S
®
ZNB/ZNBT
617
User Manual 1173.9163.02 ─ 53
Permittivity and velocity factor are coupled parameters, i.e. setting one of them deter-
mines the other. A higher permittivity implies a smaller mechanical length, but leaves
delay and electrical length unchanged (see
Max. DUT Measure Delay / Max. El.
).
Remote command:
[SENSe<Ch>:]HARMonic:PERMittivity:DATA
[SENSe<Ch>:]HARMonic:VELocity:DATA
Rise Time Definition
Selects between rise time definitions 10%-90% (default) and 20%-80%.
Remote command:
[SENSe<Ch>:]HARMonic:RTIMe:THReshold
Rise Time
The minimum rise time (according to the selected
wishes to measure on the DUT.
This in turn determines the stop frequency of the frequency sweep: the smaller the
"Rise Time", the larger the stop frequency – limited by the analyzer's maximum fre-
quency.
Remote command:
[SENSe<Ch>:]HARMonic:RTIMe:DATA
TDR Stimulus - Advanced Settings Dialog
The "TDR Stimulus - Advanced Settings " compiles advanced settings of the time
domain transformation that is available with the standard time domain option
R&S
ZNB/ZNBT-K2.
Access:
> "Advanced Settings..."
Figure 6-58: TDR Stimulus – Advanced Settings dialog
Chapter 6.5.5, "Time Domain Tab"
Applic Softtool