Command Reference
R&S
®
ZNB/ZNBT
1054
User Manual 1173.9163.02 ─ 53
Manual operation:
See
"Add / Add All... / Replace / Apply / Apply to All"
MMEMory:STORe:CORRection:TCOefficient<Ch>
<PmclFile>
Saves the power meter (two-port) transmission coefficients of channel <Ch> to a
power meter correction list file.
Suffix:
<Ch>
.
Channel number
Setting parameters:
<PmclFile>
String parameter specifying the name and directory of the cre-
ated power meter correction list file. The file extension
*.pmcl
is mandatory.
If no path is specified, the analyzer searches the current direc-
tory, to be queried with
*RST:
n/a
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:
Usage:
Setting only
Manual operation:
See
MMEMory:STORe:EYE:MASK
<TraceName>, <TraceFile>
Stores a user-defined eye mask to a 7bit ASCII file.
Setting parameters:
<TraceName>
Name of the related eye diagram
<TraceFile>
String parameter containing the path and file name of the eye
mask file. If the path is omitted, the current directory is used (see
Usage:
Setting only
Options:
R&S
ZNB/ZNBT-K20
Manual operation:
See
"Save / Load Mask Configuration"
MMEMory:STORe:EYE:MASK:RESults
<TraceName>, <TraceFile>
Saves the detailed results of the mask test in the related eye diagram to an ASCII file
(see
This command will raise an execution error if
is
OFF
.
Setting parameters:
<TraceName>
The name of the eye diagram whose mask test results shall be
exported
SCPI Command Reference