![Rohde & Schwarz 1311.6010.22 User Manual Download Page 1232](http://html.mh-extra.com/html/rohde-and-schwarz/1311-6010-22/1311-6010-22_user-manual_14777681232.webp)
Command Reference
R&S
®
ZNB/ZNBT
1232
User Manual 1173.9163.02 ─ 53
<PhyPt>
Calibrated port number. This parameter is ignored; the tolerance
value is valid for all sources.
Parameters:
<Tolerance>
Tolerance value
Range:
0.001 dB to 1000 dB
*RST:
1 dB
Default unit: dB
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]
Manual operation:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect:CFACtor
<Convergence>
Specifies the convergence factor for a source power calibration.
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the con-
vergence factor is valid for all sources.
<PhyPt>
Calibrated port number. This parameter is ignored; the conver-
gence factor is valid for all sources.
Parameters:
<Convergence>
Convergence factor
Range:
0 to 2
Increment: 0.1
*RST:
0.8
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]
Manual operation:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect:FLATness
<arg0>
Enables or disables the source power calibration. The command is valid for all chan-
nels, ports and external generators.
It is not possible to disable flatness calibration and the reference receiver calibration
simultaneously
(
SOURce<Ch>:POWer<PhyPt>:CORRection:COLLect:RRECeiver
).
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the verifi-
cation is either enabled or disabled for all channels.
<PhyPt>
Calibrated port number. This parameter is ignored; the verifica-
tion is either enabled or disabled for all sources.
SCPI Command Reference