![Rohde & Schwarz 1311.6010.22 User Manual Download Page 1112](http://html.mh-extra.com/html/rohde-and-schwarz/1311-6010-22/1311-6010-22_user-manual_14777681112.webp)
Command Reference
R&S
®
ZNB/ZNBT
1112
User Manual 1173.9163.02 ─ 53
Example:
CORR:COLL:CONN:PORTS SING
Select independent connector types at the ports.
CORR:COLL:CONN1 N50MALE; CONN4 N75FEMALE;
CONN2?
Select independent connector types at ports 1 and 4. The con-
nector type at port 2 is not changed; the query returns
N50FEMALE
.
Manual operation:
See
"Same Connector All Ports / Same Gender All Ports"
[SENSe<Ch>:]CORRection:COLLect:DELete
[<CalName>]
Deletes system error correction data generated and stored previously.
Suffix:
<Ch>
.
Channel number of the calibrated channel
Setting parameters:
<CalName>
Name of the calibration (string parameter) defined together with
the calibration type (
[SENSe<Ch>:]CORRection:COLLect:
ALL
- the analyzer deletes all calibrations.
If nothing is specified the analyzer deletes the last system error
correction stored by means of
*RST:
-
Example:
CORR:COLL:METH:DEF 'Test',RSHort,1
Select a one-port normalization at port 1 with a short standard
as calibration type.
CORR:COLL:SEL SHOR,1
Measure a short standard connected to port 1 and store the
measurement results of this standard.
CORR:COLL:SAVE:SEL
CORR:COLL:DEL ALL
Calculate the system error correction data and apply it to the
active channel, then delete the data.
Usage:
Setting only
Manual operation:
See
[SENSe<Ch>:]CORRection:COLLect:DISCard
Terminates a system error correction, discarding the acquired data.
Suffix:
<Ch>
.
Channel number of the calibrated channel
SCPI Command Reference