Command Reference
R&S
®
ZNB/ZNBT
1320
User Manual 1173.9163.02 ─ 53
Note:
If the specified cal kit does not exist, it is created with the specified calibration
standard.
Suffix:
<Ch>
.
Channel number. This suffix is ignored because calibration kits
are channel-independent.
Parameters:
<ConnType>
Connector type, one of the following identifiers:
N50, N75:
N 50 Ω or N 75 Ω connectors
PC7, PC35, PC292:
PC 7, PC 3.5 or 2.92 mm connectors
USER<no>:
User-defined connectors UserConn1, UserConn2
SMA:
User-defined connector type SMA ...
Note:
This command only supports ZVR-compatible connector
types. For general definitions use
<StandardType>
Standard type. For reflection standards, the first character
denotes the gender, e.g.:
FOPEN, MOPEN:
Open (f) or Open (m) standard.
For transmission standards, the first two characters denotes the
genders on both ends, e.g.:
FFSNetwork, MFSNetwork, MMSNetwork:
Symm. network
(ff), symm. network (mf) or symm. network (mm) standard.
For a complete list of standard types refer to
.
Parameter list
String parameters to specify the configured standard
(
<CalkitName>
,
<StandardLabel>
) and numeric parameters
defining its properties. See
.
*RST:
n/a
Example:
CORR:CKIT:N50:FOPEN 'ZV-Z21','',
0,1.8E+010,0.0151,0,0,0.22,-0.22,0.0022
Define the properties of the open (f) standard for the N 50 Ω
connector type contained in the ZV-Z21 calibration kit: Assign a
valid frequency range of 0 Hz to 18 GHz, an electrical length of
15.1 mm, 0 dB loss and define the polynomial coefficients of the
fringing capacitance as 0 fF, 0.22 fF/GHz, –0.22 fF/(GHz)
2
,
0.0022 fF/(GHz)
3
.
[SENSe:]CORRection:CKIT:INSTall
<CalKitFile>
Loads cal kit data from a specified R&S ZVR cal kit file.
R&S ZVR/ZVABT Compatible Commands