SYSTem:CORRection:WIZard[:IMMediate]
<Dialogs>
Keysight-compatible command to open the
or the
Tip:
When working with the
Chapter 7.1.3, "GPIB Explorer"
raw mode ("Options">"Raw mode" in the IECWIN32 GUI) before executing this com-
mand.
Setting parameters:
<Dialogs>
MAIN | CKIT
MAIN: open the "Calibration > Start Cal" softtool tab
CKIT: open the "Calibration Kits" dialog
Usage:
Setting only
SYSTem:DATE
<Year>, <Month>, <Day>
The command queries or defines the instrument's current date setting.
The setting command requires administrator rights; refer to
Parameters:
<Year>
Year, four-digit number
<Month>
Month, two-digit number, 01 (for January) to 12 (for December)
<Day>
Day, two-digit number, 01 to the number of days in the month
Example:
SYST:DATE?
Response:
2012,05,01
- it is the 1st of May, 2012.
SYSTem:DFPRint?
Queries the device footprint. The device footprint contains detailed information about
the instrument and is mostly used for service purposes.
Usage:
Query only
Manual operation:
See
"Save... / Print... / Save Report"
SYSTem:DISPlay:BAR:HKEY[:STATe]
<Boolean>
SYSTem:DISPlay:BAR:MENU[:STATe]
<Boolean>
SYSTem:DISPlay:BAR:STATus[:STATe]
<Boolean>
SYSTem:DISPlay:BAR:STOols[:STATe]
<Boolean>
SYSTem:DISPlay:BAR:TITLe[:STATe]
<Boolean>
SYSTem:DISPlay:BAR:TOOLs[:STATe]
<Boolean>
Displays or hides the hardkey panel (
HKEY
), the menu bar below the diagram area
(
MENU
), the status bar below the diagram area (
STATus
), the softtool panel (
STOols
),
the title bar of the main VNA application window (
TITLe
), and the toolbar above the
diagram area (
TOOLs
).
SCPI Command Reference