GUI Reference
R&S
®
ZNB/ZNBT
431
User Manual 1173.9163.02 ─ 53
Note:
Transmission normalization and "One Path Two Ports" calibration types require
two-port (Through) characterization data for the cal unit. These two-port characteriza-
tions can be unavailable in the factory characterizations of some older calibration units
or in a user characterization. If a tooltip indicates missing two-port characterization
data, simply perform a new characterization of your cal unit. In the first dialog of the
"Characterization" wizard, select "Take All OSM and Through" to make sure that the
necessary two-port data is acquired. See also
Remote command:
[SENSe<Ch>:]CORRection:COLLect:AUTO:TYPE
[SENSe<Ch>:]CORRection:COLLect:AUTO:PORTs:TYPE
[SENSe<Ch>:]CORRection:COLLect:AUTO:CONFigure
Step 2: Connections
Defines the port assignments between the R&S
ZNB/ZNBT and the calibration unit.
Background information
Refer to
Chapter 5.5.5.4, "Multiple Port Assignments"
on page 172 for details on multi-
ple port assignments.
Figure 6-17: Multiple calibrations (and multiple port assignments)
Similar functionality as for the "SingleCal" version of the wizard (see
on page 422): select the adequate calibration in the "Cal Type"/"Ports" table to
display and edit the corresponding port assignments.
Back
Go back to
Start
If the configured port assignments are invalid, this action is disabled.
Cal Softtool