![Rohde & Schwarz 1311.6010.22 User Manual Download Page 1285](http://html.mh-extra.com/html/rohde-and-schwarz/1311-6010-22/1311-6010-22_user-manual_14777681285.webp)
Command Reference
R&S
®
ZNB/ZNBT
1285
User Manual 1173.9163.02 ─ 53
SYSTem:DISPlay:TRACes:CCOunt
<ColorCount>
Defines the maximum number of trace colors (trace properties for the user defined
color scheme defined using
Parameters:
<ColorCount>
Range:
1 to 100
Manual operation:
See
SYSTem:DISPlay:UPDate
<Activate>
Switches the display on or off while the analyzer is in the remote state. The command
has no effect while the analyzer is in the local operating state.
Tip:
Switching off the display speeds up the measurement. This command may have
an impact on the update of trace and channel settings; see
.
Parameters:
<Activate>
ON | OFF - switch the display on or off. If the display is switched
on, the analyzer shows the diagrams and traces like in manual
control.
ONCE - switch the display on and show the current trace. This
parameter can be used for occasional checks of the measure-
ment results or settings. The measurement is continued, how-
ever, the measurement results are not updated. Compared to
the
ON
setting,
ONCE
does not slow down the measurement
speed.
Example:
SYST:DISP:UPD ON
Switch the display on to view the traces and diagrams.
SYSTem:ERRor[:NEXT]?
Queries and at the same time deletes the oldest entry in the error queue. Operation is
identical to that of
The entry consists of an error number and a short description of the error. Positive
error numbers are instrument-dependent. Negative error numbers are reserved by the
SCPI standard; see section
Error Messages and Troubleshooting
.
Example:
SYST:ERR?
Query the oldest entry in the error queue.
0,"No error"
is
returned if the error queue is empty.
Usage:
Query only
SYSTem:ERRor:ALL?
Queries and at the same time deletes all entries in the error queue.
SCPI Command Reference