Command Reference
R&S
®
ZNB/ZNBT
1296
User Manual 1173.9163.02 ─ 53
SYSTem:SOUNd:ALARm[:STATe]
<Boolean>
SYSTem:SOUNd:STATus[:STATe]
<Boolean>
These commands switch alarm or status sounds on or off.
Parameters:
<Boolean>
ON | OFF
*RST:
ON
Example:
SYST:SOUN:ALAR OFF; STAT OFF
Switch alarm and status sounds off.
Manual operation:
See
SYSTem:TIME
<Hours>, <Minutes>, <Seconds>
The command queries or defines the instrument's current time setting.
The setting command requires administrator rights; refer to
Parameters:
<Hours>
Range:
0...23
<Minutes>
Range:
0...59
<Seconds>
Range:
0...59
Example:
SYST:TIME?
Response:
12,0,0
- it is precisely 12 pm.
SYSTem:TRESet[:STATe]
<Boolean>
Defines the behavior of the
INITiate<Ch>[:IMMediate][:DUMMy]
.
Parameters:
<Boolean>
OFF (0)
Previous trace data are preserved.
ON (1)
Previous trace data are deleted.
This is a global setting that is not affected by an instrument
reset. Factory default is OFF (0).
Manual operation:
See
"Restart: Set all Traces to 0"
SYSTem:TSLock
<Type>
Locks the touchscreen functionality of the R&S
ZNB/ZNBT, e.g. in order to prevent
inadvertent entries during remote control.
SCPI Command Reference