
Command Reference
R&S
®
ZNB/ZNBT
1305
User Manual 1173.9163.02 ─ 53
Suffix:
<Ch>
.
Channel number
Parameters:
<Slope>
POSitive | NEGative | HIGH | LOW
Trigger slope for the triggered measurement sequence:
POSitive | NEGative - rising or falling edge
HIGH | LOW - high or low level
Example:
TRIG:SOUR EXT
Activate external signal as trigger source.
TRIG:SLOP NEG
Trigger on the negative edge of the (external TTL) trigger signal.
Manual operation:
See
TRIGger<Ch>[:SEQuence]:SOURce
<TrigSource>
Selects the source of the trigger events that the analyzer uses to start a measurement
sequence.
Suffix:
<Ch>
.
Channel number
Parameters:
<TrigSource>
IMMediate | EXTernal | MANual | MULTiple
IMMediate
Free run measurement (untriggered)
EXTernal
Trigger by external signal applied to the EXT TRIG IN connector
or pin 2 of the USER PORT on the rear panel.
MANual
Trigger event generated by pressing the "Manual Trigger" soft-
key
MULTiple
Multiple trigger mode, configured by
TRIGger<Ch>[:SEQuence]:MULTiple...
commands
Example:
TRIG:SOUR MAN
Activate manual trigger mode. The analyzer starts the next mea-
surement sequence when the "Manual Trigger" button is
pressed.
Manual operation:
See
"FreeRun / External / Manual / Multiple Triggers"
8.3.19.2
TRIGger:CHANnel:AUXiliary...
The
TRIGger:CHANnel<Ch>:AUXiliary...
commands control the external trigger
output EXT TRIG OUT at the rear panel of the instrument. The output trigger signal
SCPI Command Reference