
Getting Started
R&S
®
ZNB/ZNBT
39
User Manual 1173.9163.02 ─ 53
The USB ports can be used to connect:
●
External PC accessories such as mouse or other pointing devices, a keyboard,
printer or external storage device (USB stick, CD-ROM drive etc.).
●
External measurement equipment such as a calibration unit, power meter, signal
generator or switch matrix.
4.2.2
Front Panel R&S
ZNBT
The largest part of R&S
ZNBT front panel is reserved for the test ports, with an admin-
istrative area to the right. Brief explanations on the connectors and controls and the
rear panel can be found on the next pages.
Figure 4-3: Front view of R&S
ZNBT40 (fully equipped with the maximum 24 ports)
Instrument damage caused by cleaning agents
Cleaning agents contain substances such as solvents (thinners, acetone, etc.), acids,
bases, or other substances. Solvents can damage the front panel labeling, plastic
parts, or screens, for example.
Never use cleaning agents to clean the outside of the instrument. Use a soft, dry, lint-
free dust cloth instead.
4.2.2.1
Test Ports
Instrument Tour