GUI Reference
R&S
®
ZNB/ZNBT
422
User Manual 1173.9163.02 ─ 53
Common Port
This combo-box is only available if
"Calibration" tab of the "System Config" dialog.
It defines the port that must appear in each port assignment of an automatic calibra-
tion, and that will be used as the "center" of all Through measurements.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:AUTO:CPORt
Select All/Deselect All
Selects/deselects all ports.
These buttons are only shown if more than 4 ports are available.
Next
. "Next" is unavailable (and a warning is displayed) if
the following happens:
●
The selected characterization data do not cover all the ports to be calibrated.
●
The selected characterization data do not contain all standards needed for the
selected calibration type.
Step 2: Connections
Defines the port assignments between the R&S
ZNB/ZNBT and the calibration unit.
Background information
Refer to:
●
Chapter 5.5.5.4, "Multiple Port Assignments"
on page 172 for details on multiple
port assignments
●
on page 455 for SMARTerCal.
Depending on the number of test ports, the user interface is slightly different:
Cal Softtool