This is a global setting.
Parameters:
<SweepEndModes> AUTO | REDuce | KEEP
AUTO: at sweep end restore the power at sweep start
REDuce: at sweep end reduce the output power as if the chan-
nel base power was set to its minimum possible value
KEEP: at sweep end keep the power at its current level
Example:
SOURce:POWer:SWEepend:MODE REDuce
Reduce power at sweep end.
SOURce:POWer:SWEepend:SDELay 0.01
Use a settling delay of 10 ms
SOURce:POWer:SWEepend:MODE AUTO
At sweep end, restore the power at sweep start
Manual operation:
See
SOURce:POWer:SWEepend:SDELay
<SettlingDelay>
If sweep end mode REDuce or KEEP is active (see
) this defines the settling delay at subsequent sweep starts, i.e. the time between
power reset and sweep start when a new sweep is requested.
This is a global setting.
Parameters:
<SettlingDelay>
Settling delay
Default unit: s
Manual operation:
See
"Settling Delay / Reset Delay"
SOURce<Ch>:POWer<PhyPt>[:LEVel][:IMMediate][:AMPLitude]
<IntSourcePow>
Defines the power of the internal signal source (channel base power). The setting is
valid for all sweep types, except power sweep.
Tip:
and
to define the sweep range for a power sweep.
Suffix:
<Ch>
.
Channel number
<PhyPt>
Test port number of the analyzer. This suffix is ignored because
the selected channel base power applies to all source ports
used in the active channel.
It is possible though to define a port-specific slope factor for the
source power (
SOURce<Ch>:POWer<PhyPt>[:LEVel][:
) and a port-specific power offset
(
SOURce<Ch>:POWer<PhyPt>[:LEVel][:IMMediate]:
SCPI Command Reference