Start Power / Stop Power
Defines the sweep range for
sweeps.
Remote command:
SOURce<Ch>:POWer<PhyPt>:STARt
SOURce<Ch>:POWer<PhyPt>:STOP
Step Attenuators
Defines an attenuation factor for the received waves at each analyzer port. At least one
receiver step attenuator must be installed to use this feature; see
Receiver step attenuators are used to adapt the received signal level to the input level
range of the analyzer. They protect the instrument from being overloaded or damaged,
e.g. if the DUT is a power amplifier.
Attenuation factors are port- and channel-specific. Possible values are 0
dB to 30
dB in
steps of 10
dB. The analyzer rounds any entered value below the maximum attenua-
tion to the closest step.
Note:
Receiver step attenuators are optional hardware (see
If the number of test ports (on the VNA and connected switch matrices) is higher than
4, the "Step Attenuators" input fields are hidden. They are replaced by a "Step Attenua-
tors..." button that opens the following dialog:
Remote command:
RF Off All Channels
"RF Off All Channels" switches the internal power source for all channels off (if
checked) or on. Switching off the RF power helps to prevent overheating of a connec-
ted DUT while no measurement results are taken.