GUI Reference
R&S
®
ZNB/ZNBT
463
User Manual 1173.9163.02 ─ 53
The way the analyzer fimware activates/deactivates ports after a successfull calibration
(system error correction or power calibration) has slightly changed:
●
In all FW releases, calibrated ports that were previously disabled, are automatically
enabled as single-ended logical ports.
●
Starting with FW version 2.10, uncalibrated ports were disabled in the related
channel(s). If only one of the physical ports forming a balanced port was calibrated,
the balanced port was dissolved and only the uncalibrated (single-ended) port was
disabled.
In earlier FW versions there was no such port deactivation mechanism.
●
Since FW version 2.40 an uncalibrated port is only disabled if it is not used by a
measurement, i.e. if it is not required by any trace of the related channel.
Remote command:
The port parameters in many calibration commands define ports to be calibrated.
Cal Unit
Displays the connected calibration units. The R&S
ZNB/ZNBT auto-detects all calibra-
tion units which are connected to one of its USB ports. If several cal units are connec-
ted, one of them must be selected for calibration (active cal unit). A warning is dis-
played if the current sweep range of the network analyzer exceeds the characterized
frequency range of the calibration unit.
For background information, see
Chapter 5.5.5, "Automatic Calibration"
Remote command:
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess:ALL?
SYSTem:COMMunicate:RDEVice:AKAL:ADDRess
Characterization
Displays all characterizations that are stored in the active cal unit. The "Factory" char-
acterization is available for all calibration units; it ensures an accurate calibration for all
standard applications. To account for modifications of the cal unit such as the connec-
tion of additional adapters, you can generate modified sets of characterization data
using the cal unit characterization wizard. See
Chapter 6.11.2.3, "Characterize Cal Unit
on page 486. By default, the R&S
ZNB/ZNBT uses the last generated cal unit
characterization.
Tip:
If the characterization wizard is password-protected, the "Characterization" button
is unavailable. Use this functionality to prevent inadvertent activation of inappropriate
characterizations. See
See also
Chapter 5.5.5.3, "Characterization of Calibration Units"
Remote command:
[SENSe<Ch>:]CORRection:COLLect:AUTO:CONFigure
SYSTem:COMMunicate:RDEVice:AKAL:CKIT:CATalog?
SYSTem:COMMunicate:RDEVice:AKAL:CKIT:STANdard:CATalog?
SYSTem:COMMunicate:RDEVice:AKAL:SDATa?
Query further cal unit properties:
SYSTem:COMMunicate:RDEVice:AKAL:DATE?
SYSTem:COMMunicate:RDEVice:AKAL:FRANge?
SYSTem:COMMunicate:RDEVice:AKAL:PORTs?
SYSTem:COMMunicate:RDEVice:AKAL:WARMup[:STATe]?
Cal Softtool