Command Reference
R&S
®
ZNB/ZNBT
1173
User Manual 1173.9163.02 ─ 53
AUTO
automatic/adaptive gain control (AGC) according to the RF input
level
LDIStortion
fixed, small IF gain, for high input levels
LNOise
fixed, large IF gain, for low input levels
*RST:
LDIStortion
Example:
*RST; :SENSe:POWer:GAINcontrol:GLOBal MANual
Enable the manual configuration of independent GC settings for
each drive port.
SENSe:POWer:GAINcontrol:ALL LNO
Set the GC mode of all a and b wave receivers to "Low Noise".
Usage:
Setting only
Manual operation:
See
[SENSe<Ch>:]POWer:GAINcontrol:GLOBal
<GCModesGlobal>
Globally configures the gain control (GC) in all receive paths (measurement receivers,
b-waves) for all analyzer ports or enables port-specific gain control configuration.
Suffix:
<Ch>
.
Channel number
Parameters:
<GCModesGlobal>
LNOise | LDIStortion | AUTO | MANual
AUTO
automatic/adaptive gain control (AGC) according to the RF input
level
LDIStortion
fixed, small IF gain, for high input levels
LNOise
fixed, large IF gain, for low input levels
MANual
Enables the manual configuration of independent AGC settings
for each segment, drive port and receiver. See
*RST:
LDIS
Example:
See
Manual operation:
See
8.3.14.16
[SENSe:]ROSCillator...
The
[SENSe:]ROSCillator...
commands control the frequency reference signal.
SCPI Command Reference