GUI Reference
R&S
®
ZNB/ZNBT
677
User Manual 1173.9163.02 ─ 53
Remote command:
n.a.
Use Default Tab for Hardkey
If the checkbox is selected (system default), the
activate the first
enabled tab of their associated softtool. Otherwise the last used tab is activated.
For background information, see
Chapter 6.1, "Function Keys and Softtools"
Remote command:
n/a
Conductance in Embedding Networks
Changes the presentation of "capacitance C<i> in parallel with resistance R<i>" circuit
blocks in lumped de/embedding networks (see
Chapter 5.6.2.3, "Circuit Models for 2-
Chapter 5.6.2.4, "Circuit Models for 4-Port Net-
If active, the resistance R<i> is displayed and specified as conductance G<i> (=1/
R<i>).
Remote command:
n/a
Restart: Set all Traces to 0
Defines the behavior of the "Restart Sweep" function (see
If unchecked (factory default) the previous measurement data are preserved. If
checked, they are deleted.
Remote command:
Number of Trace Colors
Defines the maximum number of trace colors (trace properties for the
).
Remote command:
Decimal Places
Defines the number of fractional digits for quantities with different physical units. The
settings affect entries and results, e.g. the values in the marker lists.
Note:
If your instrument is equipped with option R&S
ZNB/ZNBT-K19, 1 mHz Fre-
quency Resolution, set "Decimal Places" of unit "Hz" to
12
to utilize the high frequency
resolution.
Remote command:
n/a
Setup Softtool