GUI Reference
R&S
®
ZNB/ZNBT
679
User Manual 1173.9163.02 ─ 53
Show Remote Error Info Messages
Defines whether information popups are displayed whenever a remote control com-
mand error occurs.
The displayed information can be useful for program development and optimization; it
does not necessarily indicate that a remote control script is faulty or non-executable.
Note
●
If either "Show Instrument Messages" or "Show Info Messages" is unchecked, this
button is disabled. However, its checked state is memorized.
●
For SCPI error
-113, Undefined header
, no tooltip is displayed.
Remote command:
Channel Bits Tab
Sets a channel-dependent 8-bit decimal value (0, ..., 255) to control eight independent
output signals at the USER PORT connector (lines 8 to 11 and 16 to 19).
Setting the channel bits does not change the analyzer state.
Channel Bits (Decimal)
Entry of the 8-bit decimal value (0 ... 255) for the selected channel. The channel bits
control eight output signals at the USER PORT connector. The signals are 3.3 V TTL
signals which can be used to differentiate between up to 256 independent analyzer
states. For an application example, refer to the description of the remote-control com-
mand.
The decimal values have the following effect:
●
0 means that no output signals are enabled at any of the pins 8, 9, 10, 11 and 16,
17, 18, 19.
●
1 enables the output signal at pin 8. The signal is switched on while a measure-
ment sweep is running in the selected channel. All other signals are inactive.
●
2 enables the output signal at pin 9.
●
3 enables the output signals at pins 8 and 9.
●
...
●
255 enables the output signals at all pins. See also
Remote command:
Setup Softtool