Command Reference
R&S
®
ZNB/ZNBT
815
User Manual 1173.9163.02 ─ 53
Manual operation:
See
CALCulate<Chn>:LIMit:CONTrol:SHIFt
<LimShift>
Shifts an existing limit line in horizontal direction. See also
Suffix:
<Chn>
.
Channel number used to identify the active trace
Setting parameters:
<LimShift>
Offset value for the limit line
Range:
Virtually no restriction for limit segments
Default unit: NN
Example:
*RST; :CALC:LIM:CONT 1 GHZ, 2 GHZ
Define a limit line segment in the stimulus range between 1 GHz
and 2 GHz, using default response values.
CALC:LIM:CONT:SHIF 1; :CALC:LIM:CONT?
Shift the segment by 1 Hz. The modified limit line segment
ranges from
1000000001
(Hz) to
2000000001
(Hz).
Usage:
Setting only
Manual operation:
See
CALCulate<Chn>:LIMit:DATA
<Type>, <StartStim>, <StopStim>, <StartResp>,
<StopResp>
Defines the limit line type, the stimulus and response values for a limit line with an arbi-
trary number of limit line segments. See
Chapter 5.4.1.1, "Rules for Limit Line Defini-
Note:
CALCulate<Chn>:LIMit:CONTrol[:DATA]
does not overwrite existing limit line segments. The defined segments are appended to
the segment list as new segments.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Type>
Identifier for the type of the limit line segment:
0
–
limit line segment off, segment defined but no limit check
performed.
1
–
upper limit line segment
2
–
lower limit line segment
Range:
0, 1, 2 (see above)
SCPI Command Reference