Command Reference
R&S
®
ZNB/ZNBT
1235
User Manual 1173.9163.02 ─ 53
SOURce<Ch>:POWer<PhyPt>:CORRection:DATA:PARameter<Wv>:COUNt?
Gets the number of available power calibrations (both source power and measurement
receiver).
Use
SOURce<Ch>:POWer:CORRection:DATA:PARameter<Wv>?
to retrieve the
related settings.
Suffix:
<Ch>
.
Number of the calibrated channel.
<PhyPt>
This suffix is ignored.
<Wv>
This suffix is ignored.
Usage:
Query only
Manual operation:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:GENerator<Gen>[:STATe]
<Boolean>
Enables or disables the source power calibration for channel
<Ch>
and for an external
generator number
<Gen>
. The command is disabled unless a source power calibration
for the external generator has been performed (
). To enable or disable a source power calibration for an
SOURce<Ch>:POWer<PhyPt>:CORRection:STATe
.
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt>
Calibrated port number. This suffix is ignored; the generator is
selected via
<Gen>
.
<Gen>
Generator number
Parameters:
<Boolean>
Enables (ON) or disables (OFF) the source power calibration for
port number
<PhyPt>
.
*RST:
OFF
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]
Manual operation:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:IMODulation:LTONe[:ACQuire]
Starts the source calibration for the lower tone (1
st
power calibration step for intermo-
dulation measurements), stores and applies the calibration data. The external power
meter used is selected via
SOURce<Ch>:POWer<PhyPt>:CORRection:PMETer:ID
.
Suffix:
<Ch>
.
Calibrated channel number
SCPI Command Reference