GUI Reference
R&S
®
ZNB/ZNBT
468
User Manual 1173.9163.02 ─ 53
Ports
Selects the test ports to be calibrated.
It is possible to select any combination of two or more test ports. If you are only inter-
ested in a single port
p
, perform a two-port SMARTerCal for a port pair including
p
(as
Power Port).
Note:
Calibration and port de-/activation.
The way the analyzer fimware activates/deactivates ports after a successfull calibration
(system error correction or power calibration) has slightly changed:
●
In all FW releases, calibrated ports that were previously disabled, are automatically
enabled as single-ended logical ports.
●
Starting with FW version 2.10, uncalibrated ports were disabled in the related
channel(s). If only one of the physical ports forming a balanced port was calibrated,
the balanced port was dissolved and only the uncalibrated (single-ended) port was
disabled.
In earlier FW versions there was no such port deactivation mechanism.
●
Since FW version 2.40 an uncalibrated port is only disabled if it is not used by a
measurement, i.e. if it is not required by any trace of the related channel.
Remote command:
The port parameters in many calibration commands define ports to be calibrated.
Type
Selects the calibration type to be performed. For background information, see
ter 5.5.7.2, "Calibration Types"
SMARTerCal is based on a full n-port calibration (n > 1): "Adapter Removal" is a 2-port
calibration, all other calibration types can be used for any combination of 2 or more
ports.
Remote command:
[SENSe<Ch>:]CORRection:COLLect:METHod:DEFine
Power Meter
Shows a list of all power meters that have been properly configured and are available
for the scalar source power calibration of a single source port. The last configured
power meter is selected by default. See
Remote command:
[SENSe:]CORRection:COLLect:PMETer:ID
Power Port
Selects the port for the scalar source power calibration.
Select All/Deselect All
Selects/deselects all ports.
These buttons are only shown if more than 4 ports are available.
Next
"Step 2: Connectors and Cal Kits"
on page 469. This button is inactive
(and a warning is displayed) if fewer than two ports are selected.
Cal Softtool