GUI Reference
R&S
®
ZNB/ZNBT
680
User Manual 1173.9163.02 ─ 53
Pin 16 - 19
Selects the control mechanism for the signals at pins 16 to 19 of the USER PORT con-
nector.
●
"Channel Bits":
Signals are controlled by channel bits 4 to 7. No drive port indica-
tion at the USER PORT connector.
●
"Drive Ports":
Signals indicate the active drive ports. The number of active chan-
nel bits is reduced to 4 (pins 8 to 11).
Remote command:
Advanced Tab
Collects several advanced settings.
Geometric Calculation of Bandfilter Center
Defines how bandfilter searches calculate the center frequency of the passband or
stopband (see
If "Geometric Calculation of Bandfilter Center" is checked, the
geometric mean
of the
lower band edge and upper band edge frequencies is used, otherwise their
arithmetic
mean
.
Remote command:
CALCulate:MARKer:FUNCtion:BWIDth:GMCenter
Error Dump Type
Determines the level of detail ("Minimal", "Normal", "Large", "Full") and hence the size
of the dump files that are created if a firmware exception occurs. "None" disables dump
file creation.
The latest 5 exception dumps can be retrieved using the "Save Report" function of the
"Info" dialog (see
"Save... / Print... / Save Report"
Chapter 10, "Error Messages and Troubleshooting"
Remote command:
Setup Softtool