Concepts and Features
R&S
®
ZNB/ZNBT
122
User Manual 1173.9163.02 ─ 53
U
0
I
1
I
2
U
1
U
2
U
0
Z
0j
Z
ij
Z
0i
The calculation formula of a converted serial transmission impedance Z
ij
depends on
the waveguide circuit theory according to which
Table 5-5: Calculation of Converted Series Transmission Impedances
Traveling Waves
Power Waves
Parallel Transmission Impedance
A two-port transmission parameter Z
ij
(i ≠ j) can also describe a parallel impedance
between the two ports.
U
0
I
1
I
2
U
1
U
2
U
0
Z
0j
Z
ij
Z
0i
The calculation formula of a converted parallel transmission impedance Z
ij
depends on
the waveguide circuit theory according to which
Table 5-6: Calculation of Converted Parallel Transmission Impedances
Traveling Waves
Power Waves
Shunt-thru Measurements
The shunt-thru method is used for measuring very low impedances. A typical applica-
tion are measurements on power distribution network (PDN) components, such as
bypass capacitors and DC-DC converters.
Measurement Results