Command Reference
R&S
®
ZNB/ZNBT
1239
User Manual 1173.9163.02 ─ 53
SOURce<Ch>:POWer<PhyPt>:CORRection:STATe
<Boolean>
Enables or disables the source power calibration for channel
<Ch>
and for port number
<PhyPt>
. The setting command is disabled unless a source power calibration for the
analyzer port has been performed (
SOURce<Ch>:POWer<PhyPt>:CORRection[:
). The query always returns a result.
To enable or disable a source power calibration for an external generator use
SOURce<Ch>:POWer<PhyPt>:CORRection:GENerator<Gen>[:STATe]
.
Suffix:
<Ch>
.
Calibrated channel number
<PhyPt>
Calibrated port number
Parameters:
<Boolean>
Enables (ON) or disables (OFF) the source power calibration for
port number
<PhyPt>
.
*RST:
OFF
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection[:ACQuire]
.
Manual operation:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:TCOefficient[:STATe]
<Boolean>
Enables or disables the use of two-port transmission coefficients.
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the trans-
mission coefficient settings are valid for all channels.
<PhyPt>
Calibrated port number. This parameter is ignored; the transmis-
sion coefficient settings are valid for all sources.
Parameters:
<Boolean>
Enables (ON) or disables (OFF) the two-port transmission coeffi-
cients.
*RST:
n/a
Example:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:
Manual operation:
See
SOURce<Ch>:POWer<PhyPt>:CORRection:TCOefficient:CALibration
<Boolean>
Selects the position of the additional two-port in the test setup.
Suffix:
<Ch>
.
Calibrated channel number. This parameter is ignored; the trans-
mission coefficient settings are valid for all channels.
SCPI Command Reference