![Rohde & Schwarz 1311.6010.22 User Manual Download Page 136](http://html.mh-extra.com/html/rohde-and-schwarz/1311-6010-22/1311-6010-22_user-manual_1477768136.webp)
Concepts and Features
R&S
®
ZNB/ZNBT
136
User Manual 1173.9163.02 ─ 53
When the sweep axis is changed from linear frequency sweep to logarithmic sweeps,
straight limit lines are transformed into exponential curves. The sweep points are redis-
tributed along the x-axis, so the number of failed points can change.
Logarithmic interpolation
The analyzer offers a logarithmic interpolation mode that allows you to carry over the
limit line definition to logarithmic sweeps.
While "Show Limit Line" is active, the diagrams display all limit line segments.
Exception:
In a segmented frequency sweep with point-based x-axis, gaps between
the segments are minimized. To facilitate the interpretation, the R&S
ZNB/ZNBT dis-
plays only the limit line segments which provide the limit check criterion (the "tighter"
limit line at each point). In the example below, this rule results in a single, continuous
lower limit line.
Operations on Traces